Maintaining an optical fibre fusion splicer: 7 good habits for reliable splices

Electrodes, V-grooves, lenses, cleaver, arc test: the simple actions that preserve the quality of your splices and extend the life of your fusion splicer.

An optical fibre fusion splicer is a precision instrument. Well maintained, it produces very low-loss splices for years. Neglected, it leads to more rework, high losses and breakdowns at the worst moment on site. Here are the habits to adopt every day.

1. Keep the V-grooves perfectly clean

The V-grooves hold and align the fibres. The slightest residue of coating or dust shifts the fibre and degrades alignment. Clean them regularly with a cotton swab soaked in isopropyl alcohol, never using a metal object.

2. Watch the lenses and mirrors

The fusion splicer “sees” the fibres through its cameras. Dirty lenses or mirrors distort the image analysis and the loss estimate. Gentle cleaning, with the accessories recommended by the manufacturer, is enough in most cases.

3. Replace the electrodes in good time

Electrodes wear down with every arc. Worn or dirty, they produce an unstable arc and lower-quality splices. Keep track of your device’s arc counter and observe the replacement interval recommended by the manufacturer.

4. Run an arc test when conditions change

Altitude, temperature and humidity influence the electric arc. An arc test (or arc calibration) after a change of site, weather conditions or electrodes allows the fusion splicer to adjust its power and keep splices consistent.

5. Don’t forget the cleaver

A good splice starts with a good cut. A worn blade produces irregular cleave angles that increase losses. Rotate the blade according to the recommendations, keep the device clean and regularly empty the fibre scrap collector.

6. Take care over transport and storage

  • Keep the fusion splicer in its carrying case, protected from shocks and dust.
  • Avoid sudden temperature variations and humidity; let the device acclimatise before use.
  • Look after the battery: no prolonged full discharge, and no storage in direct sunlight.

7. Schedule a regular service

Even with impeccable daily maintenance, a service by a maintenance centre makes it possible to check the alignment, the arc, the cameras and the motors, to update the software and to detect wear before it becomes a breakdown.

A regularly serviced fusion splicer means less rework in the field and interventions that finish on time.

Our maintenance centre handles the diagnosis, calibration, repair and updating of your fusion splicers, cleavers and OTDRs. Submit a request: we will let you know the procedure and an intervention lead time.

A project, a question?